LABCO Brand Water Bath Shaker with the following specifications:
The water bath shaker is required for the holding & shaking of 20 litre (02 nos) or 10 litre (04 nos) borosilicate glass bottles to inactivate pertussis antigen at 60°C. This water bath shaker should comply norms to install inside cGMP compliant facility.
- Dimension:
- External (Lx W x H): 106 x 73 x 43 cm
- Inner cabinet (Lx W x H): 68 x 64 x 42 cm
- Removable bottle holder: (Lx Wx H): 57 × 57 x 37 cm
- Diameter of glass bottle holder clamp: 20L : 310 mm & 10 L : 240 mm.
- Diameter of glass bottles: 20L : 300 mm & 10 L : 230 mm.
- No. of Bottle holding Clamps: 20 L = 2 nos. OR 10 L = 4 nos.
- Maximum Bath volume: 180 L
- Shaking platform is able to bear the weight (100 Kg) and shake bottles (4 × 25 Kg).
- Temperature range: Ambient (+5°C) to 90°C.
- Temperature Uniformity: ±0.1°C
- Temperature Accuracy: ±0.1°C
- Temperature Increment: ±0.5°C
- MOC:
- Outer: Powder coated MS
- Inner: SS 304
- Lid cover: SS304 with rubber gasket
- High grade insulation (Glass wool) on all sides for minimal temperature loss
- Nonvolatile material
- Heat up time: Rapid with two booster heating coils installed in the bottom.
- Display & Control: Microprocessor based digital control digital display via soft touch keypad.
- Orbital Shaking Speed: Adjustable 30 – 100 rpm with low noise and minimal vibrations
- Orbital shaking speed deviation: ±2% of set RPM. Shaking increment: 01 RPM
- Motor: Brushless DC with stepless regulator fitted at the bottom of the shaking platform. Amplitude: 40±10 mm
- Movement of the shaker is Triple Eccentric Counter balanced drive
- We are an ISO 9001 : 2015 company
- Water Drain: At the bottom side with closed valve.
- Safety: Over-current cut off; restart of pre-set speed after power failure.
- The Water Bath shaker is equipped with heavy duty lockable wheels and anti-vibration rubber pads for immobilization.
- Power supply: 230±10 V; 50 – 60 Hz single phase.
- Warranty: 03 Years.
- IQ, OQ and PQ certificate will be provided at the time of installation.